Agilent E5250A User’s Guide, Edition 11

3-5

Executing Self-Test and Leak Test

Executing Self-Test

NOTE

If Fail LED is ON after Relay Test

After executing the relay test, if the Fail LED turns on, the cause is one of the
following:

• The Relay Test Adapter is not connected or a cable is still connected to input

connector.

• Output is not open. See step 2 of previous procedure.

• You did not properly install the E5255A. See “To Configure E5255A” on page

2-10.

• A plug-in card may be defective. Contact your nearest Agilent Technologies

Service Center.

Figure 3-1

Relay Test Adapter Connection

NOTE

Settings after Relay Test

After Relay Test, the E5250A settings are changed to same as after :SYST:CPON
command execution.

Relay Test Adapter