Agilent E5250A User’s Guide, Edition 11

11-13

Error Messages

E5250A Specific Error Messages

3013

Cannot connect multiple channels in SROUTe mode

Incorrect channel list for Single connection rule. For Single
connection rule, an input port cannot be connected to multiple output
ports, and an output port cannot be connected to multiple input ports.
Change channel list or change connection rule to Free.

3014

Cannot directly specify Bias Port channel

Incorrect channel number when Bias Mode ON is specified in the
channel list. Cannot directly specify the channel number for the Bias
Port.

3015

Bad channel number combination on E5252A card

Wrong input port combination is specified in the channel list for the
E5252A card. For input ports, can specify only one of input port
numbers 5, 7, and 9, and only one of input port numbers 6, 8, and 10.
For example, channel list cannot specify input ports 5 and 7.

3016

Config error. Re-install E5255A.

May be problem with the E5255A installation. Re-install the
E5255A.

3017

Too many relays closed. Max 200.

Too many physical relays have been closed. To close new relays,
must open some relays. Maximum 200 relays can be closed.

3018

Can't change to ACONfig mode. Check card config.

The present card configuration of the E5250A does not allow the
Auto Config mode. Change the E5250A card configuration to allow
Auto configuration mode.

3019

Cannot use same port for Couple and Bias

Same input port number is specified to be Couple Port and Bias Port.
Couple Port and Bias Port cannot be the same port.