100

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

Chapter 3

Loading and Unloading

8.

Click OK when finished.

Use the descriptions in this table to help you configure the parameters in the
Advanced Settings dialog.

Table 31 - Data Collector Advanced Settings Parameters

Parameter Name

Description

Values/Comments

Operating System

Sets the operating system version of the data collector.

2CHEnable

Enables Emonitor to download measurements configured for 2-
channel collection.

Options: Yes
No

Current Baud

The baud rate for the serial communication.

Recommend setting this to 115200.

Current Parity

The combination of parity, data bits and stop bits. The first
letter in the setting is the parity: None, Even, Odd. The first
number is the number of data bits. The second number is the
number of stop bits. N81 stands for no parity, 8 data bits, and 1
stop bit.

The Enpac uses only N81.

Datapac Enumeration

Configures the Enpac 2500 to collect data the same as the
dataPAC.

Options: Yes
No

Direction

The default direction to be used for off route measurements.
0 - None
1 - Horizontal
2 - Vertical
3 - Axial

Options:
0
1
2
3

Log Display

Display plots in log format in the data collector.

Options:
Yes
No

Machine Speed

The default machine speed to use for off route measurements.

Phone Number

Not Used

Position

The default position label to use for off route measurements.

1 to 16 characters

Sync Trigger Level

The amplitude level for externally triggering the measurement. This parameter is used when you set Trig Level to Fixed in

the Instrument Setup screen.

Refer to Configure the Data Collection Options on page 42

.

Sync Trigger Slope

The input signal slope to be used for externally triggering the
measurement.

Options:
Positive
Negative
This parameter is used when you set Trig Level to Fixed in
the Instrument Setup screen.

Refer to Configure the Data Collection Options on page 42

.

Sync Trigger Type

Sets the trigger type.

Options:
TTL
External

Trigger Range

The maximum range (in Volts) that the trigger signal can
produce.