Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

41

The Enpac 2500 Data Collector

Chapter 1

Table 7 - Setup Screen Parameter Descriptions

Parameter

Description

Value

Route Mem

Sets the location where the loaded route data is stored.

Internal (default)

Storage Card

Timeout

The number of minutes the Enpac 2500 can remain idle before it shuts itself off to
conserve battery power.

Press On/Off to reactivate the instrument following a timeout. The instrument returns to
the point in the program where it was when it shut off. No stored data will be lost.

Off (default)

5 min

10 min

15 min

20 min

Trigger

Sets the Phase/Order input trigger synchronization:

Laser Tach enables the internal Laser Tachometer when a Phase or Order

measurement is collected.

External uses the trigger input from the POWER/USB/TRIGGER connector as the

trigger source.

RS-232 transfers data can also be transferred between the data collector and your

computer.

External (default)

Laser Tach

RS-232

Dflt Units

Defines the units of measurement for Offroute standard velocity measurements.

Important: Used only* for Offroute standard velocity measurements.

English (ips) (default)

Metric (mm/s)

Date Format

The format of the date displayed in the Enpac 2500.

MM/DD/YYYY (default)

DD/MM/YYYY

YYYY/MM/DD

Time Zone

Sets the time zone.

Only used when using ActiveSync software data synchronization. The time zone on the
instrument must match the time zone on the computer. If you are not using ActiveSync
software then the time zone setting has no effect.

Important: Always set your instrument to the correct time zone or your files will not be
in sync by time zone.

List of 30 time zones

Date/Time

Displays the date and time in the format you select in the Date and Time parameters.

Current data and time

Memory

Configures the memory:

Select cmd to exit this field without selecting an option.
Delete all routes and data from memory.
Initialize formats the memory. Deletes all files from memory including any route files

and directories.

Important: These commands are performed on the memory location as set by Route
Memory.

Select cmd (default)

Delete all routes

Initialize

Route Font

You can select the font size for the Route.

Big or Small

Envelope

The type of envelope to use for high frequency and rolling element bearing analysis.

Determines which pre-defined measurement you can use to collect Offroute data.

gSE (default), collects an Offroute gSE point.

ESP, collects an Offroute ESP point

Navigation

Determines how the instrument navigates through the routes in the Data Collection
screens when the Right arrow key is pressed:

Enpac- The instrument displays a Selection screen from which you select the item

you want to collect or review.

DataPac - The instrument functions like a dataPAC 1500 in that it automatically

advances to the next point in the route.

Enpac

DataPac (default)

Transducer Change

Notification of a transducer change.

If set to Warn, a prompt appears on the screen warning you of the transducer change.

Warn (default)

Do not warn