Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 298

Advertising
background image

298

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

Chapter 11

FRF Frequency Response Function Extension Module

3.

Select an accelerometer of correct mass and connect to CH2.

4.

Place the accelerometer on mass.

The mass can be hand-held. The use of foam or a supporting structure is
not necessary.

5.

Set the Measurement parameter to Apparent Mass.

6.

Set the Meas Units parameter to N/m/s2.

The instrument displays 1 kg (or other chosen mass), plus the weight of the
accelerometer:

M = calibrated mass + accelerometer mass
For example: 1 kg + 10.5 g = 1.0105 kg

Advertising
This manual is related to the following products: