Chapter 11

FRF Frequency Response Function Extension Module

FRF Analysis Set-up Parameters

This table describes the FRF set-up parameters.

Table 54 - FRF Analysis Set-up Parameters

Parameter

Description

Value/Comments

Acquisition Mode

Set to Hammer for FRF acquisition and modal analysis, or set to
Continuous for ODS analysis.
If you are performing ODS analysis, refer to the instructions that
accompanied your ODS software for additional information on
settings that are required.

Hammer (default)
ODS analysis

Hammer Units

This setting does not restrict the Meas Units options, for example, if
Hammer Units is lbf, mobility can still be measured in mm/s/N.

N (metric default)
lbf (EU)

H.Sens

Enter the sensitivity of the hammer.

5.00 (default)
mV/N
Vlbf

Sensor Type

Specifies the type of sensor used on Connector B, for example, an
accelerometer, velocity transducer, or displacement probe.
ODS analysis requires a sensor on both Connector A and B. The
sensor type and sensitivity must match on both channels.

Accel G (default)
m/s

2

in/s
mm/s
mils
um

Sens.

Enter transducer sensitivity in mV per sensor unit.

100.00 (default)
mV/g

Measurement Sets

FRF

display type.

Mobility (default)
Apparent Mass
Accelerance
Impedance
Stiffness
Compliance

Meas Units

Measurement units change to correspond with the measurement
display type that has been chosen, but can be set to Imperial or
metric units.

ips/N (default)
mm/s/N
ips/lbf
mm/s/lbf

Y-axis

Sets Y axis to linear, logarithmic, or decibel scale.

Linear
Log (default)
dB

Hammer Range

Sets the hammer input range manually to selected fixed range, or
to be set automatically.

Automatic (default)
0.100…220000N

Response Range

Sets response sensor input range manually to selected fixed range,
or to be set automatically.
If Hammer Range or Response Range is set to Automatic, you are
prompted to make three trial hits to establish the range to use.

Automatic (default)
0.100…2000g

Freq Range

Sets the frequency range for the test. The Acquisition time
information box changes accordingly.

1000 (default)