Balancing Application

Chapter 8

6.

When the speed, vibration, and phase values are stable, press READ/OK.
This completes the plane 1 measurement.

7.

The data collector asks if the transducer is attached to plane 2. If the
transducer is attached, press F4 (OK) to begin the plane 2 measurement.

8.

The data collector automatically measures the speed, vibration, and phase
in the second plane. The values are constantly updated on the screen.
Allow these values to stabilize before continuing to the next step.

9.

Press READ/OK to complete the initial measurement for plane 1 and
plane 2. The results are stored in the Vibration Summary Table.

10.

Shut down the machine and continue the balancing procedure in the next
section.

TIP

The green status indicator illuminates when the measurement is
complete.

Refer to Status Indicators on page 25

.

IMPORTANT

Anytime during the initial run, you can do the following:

Press F4 (Esc) to exit the balancing program.
Press the 0 (Shift) key and F1 (Go to) simultaneously to go back to a

previous step in the balance run.

Refer to Moving Around in a Balance Run on page 233

for details.

Press F2 (Summary) to display the results in the Vibration Summary

Table.

Refer to Reviewing Data on page 233

for details.