Setting Up Measurements

Chapter 2

Typical Magnitude and Phase Measurement Definitions

this is what the Measurement Definition pane for one through four orders might
look like. All measurement definitions have the same units and collection
specification.

Figure 15 - Magnitude and Phase Measurement Definitions

The following table shows the collection specification for Mag & Phase.

The transducer specification in the Mag & Phase collection specification is
shown in the table below.

The filter specifications are shown in the table below.

Table 20 - Collection Specification for Mag & Phase Measurement Definitions

Transducer

Window

Signal Detection

Fmax

Lines

Phase

Order Norm?

Averages

Accelerometer

Hanning

Peak

500 Hz

400

Yes

Yes

4 linear

Table 21 - Transducer Specification in Mag & Phase Collection Specification

Name

Base Unit

Input Type

Units

Calibration Value

DC Offset

Accelerometer

Acceleration

ICP Accel

g

100

0

Table 22 - Filter Specifications in Mag & Phase Measurement Definitions

Name

Low Cutoff

High Cutoff

Data Collector Filter Type

1st Order

0.8 Orders

1.2 Orders

None

2nd Order

1.8 Orders

2.2 Orders

None

3rd Order

2.8 Orders

3.2 Orders

None

4th Order

3.8 Orders

4.2 Orders

None