Balancing Application

Chapter 8

Advanced Setup Screen

The Advanced Setup screen appears when you press F2 (Options) on the Balance
Setup screen. The Advanced Setup screen is used to set up the measurement and
enter the sensitivity of the transducer.

To edit an option in this screen, follow these steps.

1.

Highlight the option and press the RIGHT ARROW key to open a menu
of choices.

2.

Select the choice by pressing the arrow keys or type in a value using the
numeric key pad.

3.

Press the LEFT ARROW key to save your selection.

Use the descriptions in

Table 43

to help you configure the parameters in

the Advanced Setup screen.

4.

When you are finished, press F3 (Apply) to return to Balance Setup screen.

Table 43 - Advanced Setup Parameters

Parameter Name

Description

Values/Comments

Filter

Sets the high pass filter to apply to the measurement. The high pass
filter is useful in removing high vibration, low frequency signal
components, which can dominate the signal.

Options:
Off
2 Hz (default)
10 Hz
70 Hz
User Defn

Detection

Selects the type of signal detection for the measurement.
Peak - The Enpac 2500 measures the dynamic signal from zero to

peak voltage. Use this or the RMS setting for most acceleration and
velocity measurements.

Pk-Pk - The Enpac 2500 measures the dynamic signal from the

minimum peak to the maximum peak. Use this setting for most
displacement measurements.

RMS - The Enpac 2500 measures the dynamic signal as the square

root of the mean of the square. Use this or Peak for most
acceleration and velocity measurements. Also use this for voltage
measurements.

Options:
Peak
Pk-Pk
RMS

Defaults to `Peak‘ for Accel and Vel and `Pk-Pk‘ for
Disp.