Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

131

Collecting and Reviewing Data

Chapter 4

A typical spectrum could have 400 lines although the data collector lets you to
select a smaller or larger number of lines. The higher the lines, the higher the
resolution, and the longer it takes the data collector to acquire the data for a
spectrum.

Lines of resolution, Fmax, and number of averages all affect the length of time it
takes to acquire and process a spectrum. You can control each of these setup items
either in your computer software for route measurements or on the User screen
for Offroute measurements.

See

Create and Collect a User-defined Point on page 141

for more

information about Offroute measurements.

Once you have created a route in the Emonitor Software and uploaded it to the
data collector you can begin to take the measurement.

See the Emonitor User Manual, publication

EMONTR-UM001

for

information about creating routes and loading them into the data
collector.

Follow these steps to take a spectrum measurement.

1.

Attach the transducer to the current location.

2.

From the Data Collection screen, press Enter.

The Route Manager screen appears.

3.

Select the spectrum route you uploaded from the Emonitor software and
press Enter to start your measurement for the current point.

Table 33 - Tasks Associated with Ranging the Measurement

If you

Then

Previously collected a magnitude measurement for the
point

The spectrum measurement starts automatically after you
press Enter to accept the magnitude reading

Set Auto Advance to On in the Instrument Configuration
screen

The data collector automatically goes to the next point in
the list after it finishes collecting the spectrum

Set Auto Advance to Off

Then the Signature Measurement screen appears

Set Fast Averaging to On in the Instrument Configuration
screen

The data collector collects the specified number of
averages before displaying the spectrum on the screen

Set Fast Averaging to Off

The data collector displays the spectrum while collecting
the averages