The Enpac 2500 Data Collector

Chapter 1

Status Indicators

The Enpac 2500 has three status indicators located in the upper-right side of the
Enpac nameplate.

Figure 4 - Status Indicators

Red

Amber

Green

Table 4 - Status Indicators

Status Indicators

Operation

State

Definition

Red

Charging Battery

Flashing

When first connected to the power adapter, the Enpac 2500 tests the
condition of the main battery. Within 30 seconds, the status indicators
should go to a solid state. If the status indicators continues to flash, the unit
has diagnosed a fault in the main battery and it will not charge the battery.
The problem may be that the cell temperature is too high or there is a fault
with the battery.

Solid

The main battery is OK. The power adapter is connected to the Enpac 2500
and charging the main battery. The battery will be maintained by a trickle
charge from the power adapter for as long as it remains attached to the unit.

Data Collection

Solid

An alarm condition has been set.
Input signal is out of range.
ICP Fail is active.
The power adapter is connected to the Enpac 2500.

Amber

Data Collection

Solid

Acquisition sub-system is settling.

Green

Data Collection

Solid

Input signal is stable and data is ready to be stored.