Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 45

Advertising
background image

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

45

The Enpac 2500 Data Collector

Chapter 1

Acquisition

Determines how the Enpac 2500 collects a signature when the measurement definition specifies just one average.

Autoranging

Sets the input range during data collection.

If Autoranging is set to Fixed or Aggressive, the software integration is used on all FFT measurements. This is to give improved dynamic range and

faster hardware settling times.

If hardware integration is required, set Autoranging to Conservative.
For all autoranging settings (including Fixed), the input range is automatically increased and the data retaken if clipping occurs. This means that

the final input range used will never be too low for the signal being measured.

Table 8 - Instrument Configuration Parameters

Parameter

Description

Value

Value

Description

Single shot

Measures a single average for the signature and then stops.

Continuous

Default

Continuously measures the signature until you press Enter to stop the averaging. This will pause the acquisition. Press
the Continue function to restart the acquisition. Press Enter a second time to store the data.

Value

Description

Aggressive

Default

Limits the auto-ranging to 20 mV level and sets the range limits from 1% to 70% of range.

When the input is under-range, the instrument recalculates the most appropriate setting and begin data sampling. The
range will not be adjusted a second time or checked prior to data collection.

Use Aggressive autoranging when faster data collection is desired. When using Aggressive autoranging, set Auto Store to
Off, or set Auto Store Time to 2 seconds when Auto Store is set to On.

Conservative

Allows auto-ranging to go as low as 2 mV and sets the range limits from 10% to 70% of range.

The instrument perform a series of checks prior to sampling data. This will result in longer data collection times and
potentially lengthy acquisition.

When the input is under-range, the instrument recalculates the most appropriate range setting and sample data to
make sure the range has been properly selected. In addition, the dynamic range will be adjusted several times prior to
data collection.

Use Conservative autoranging when large variations in signal amplitude are present for proper input range selection but
increases the settling time required.

Fixed

No autoranging measurements are taken. Input range is set based on the transducer type and sensitivity.

For example, an accelerometer, the input range will be 5 V for 100 mV/G sensitivity or 500 mV for 10 mV/G sensitivity.

Transducer

Input Range (EUs)

Accelerometer

50 g

Velocimeter

50 ips

Displacement probe

200 mils

Volts, A.C.

2 V

Advertising
This manual is related to the following products: