Collecting run up/coast down measurements, Run up/coast down status indicators, Collecting and analyzing run up/coast down – Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 259

Advertising
background image

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

259

Run Up/Coast Down Application

Chapter 10

Collecting Run Up/Coast
Down Measurements

This section describes the Run Up/Coast Down status indicators and how to
collect Run Up/Coast Down data.

Run Up/Coast Down Status Indicators

This table describes the status indicators in the Run Up/Coast Down
application.

Collecting and Analyzing Run Up/Coast Down

When the setup parameters are complete, you can begin collecting data. The
collected data is automatically recorded as a .wav file, which can then be analyzed
using several different types of plots.

Start Spd (RPM)

Sets the speed at which the unit should start collecting data.

10…3600 RPM
Default value is 60 RPM
If Mode is set to `Runup,‘ this value must be less
than the Stop Spd. If Mode is set to `Coast down,‘
this value must be greater than the Stop Spd.
This parameter is not available if Mode is set to
`Manual.‘

Stop Spd (RPM)

Sets the speed at which the unit should stop collecting data.

10 to 3600 RPM
Default value is 300 RPM
If Mode is set to `Runup,‘ this value must be
greater than the Start Spd. If Mode is set to `Coast
down,‘ this value must be less than Start Spd.
This parameter is not available if Mode is set to
`Manual.‘

Save To

Sets the location where data is saved. We recommend that you save
the data to the SD or PC card if the .wav files will be large. Your
selection affects the maximum recording time.

Options:
Internal (default)
PC Card
SD Card

Max Acq Time (s)

The number of seconds to record the time waveform.
This limit applies only if the run up or coast down takes longer than
the time set here.

1…1131.65 seconds
Default value is 1131.65 seconds

Table 44 - Run Up/Coast Down Setup Parameters

Parameter Name

Description

Values/Comments

Table 45 - RUCD Status Indicators

Status Indicators

State

Definition

Red

Solid

An overload error (clipping or streaming error) has occurred during
data collection.

Amber

Solid

An error has occurred during data collection The status indicator
turns off when the error condition disappears.

Green

Flashing Green

No errors occurred during data collection.

Advertising