Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 80

Advertising
background image

80

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

Chapter 2

Setting Up Measurements

Set up a Collection Specification with Transducer set to the AC
displacement and Signal detection set to Peak-peak.

Finish setting up the dynamic proximity probe measurement by setting
Data Type to Magnitude and Units to mils or um. You can use a Filter if
desired.

Static (DC) component or gap

Set up a displacement transducer in the Transducer Calibration dialog
with Input type set to ‘DC Coupled’ and Units set to mils or um. Make
sure you enter the Calibration value that converts the voltage to a
displacement. If you are using the DC Input Module to remove the AC
component and attenuate the signal be a factor of 2.5, then you must
divide your transducers calibration by 2.5.

Set up a Collection Specification with Transducer set to the DC
displacement and Signal detection set to ‘RMS.’

Finish setting up the dynamic proximity probe measurement by setting
Data Type to ‘Magnitude and Units to mils or um.’ You must set the
Filter to ‘None’ since any type of high pass filtering removes the DC
component.

Typical Magnitude Measurement Definitions

This figure shows the setup for a typical magnitude measurement definition

Figure 13 - Magnitude Measurement Definition

This table shows the STD (Hz) 2000 collection specification.

This table shows the transducer specification in the STD (Hz) 2000 collection
specification.

This figure shows the setup for a typical set of measurement definitions with
multiple frequency bands. Note that the filter specifications used in the tables are
not currently defined in the Emonitor database. To define a new filter
specification, choose Setup > Band Filters and click New.

Table 17 - Collection Specification for Magnitude Measurement Definition

Transducer

Window

Signal Detection

Fmax

Lines

Phase

Order Norm?

Averages

Accelerometer

Hanning

Peak

2 kHz

400

No

No

4 linear

Table 18 - Transducer Specification in STD (Hz) 2000 Collection Specification

Name

Base Unit

Input Type

Units

Calibration Value

DC Offset

Accelerometer

acceleration

ICP Accel

g

100

0