Chapter 11

FRF Frequency Response Function Extension Module

The status indicators provide feedback on the hammer hits.

See

Status Indicators on page 25

for more information about the status

indicators:

If the Accept/Reject option in FRF - Setup was set to Automatic, each
hammer hit is collected and the Enpac 2500 data collector automatically
determines if the data is correct. If it is not, a message appears with details
on why the hit was rejected.

If Accept/Reject was set to Manual, the hammer and response time
waveforms display for both the hammer and accelerometer outputs, and
then an Accept/Reject prompt appears.

4.

Review the FRF data and either Accept or Reject the hit.

When all the averages are taken, the Taking Data screen updates to show
the FRF magnitude/phase traces.

Table 55 - Status Indicators

Status Indicator

Description

Red

An error, for example, input overrange, during data acquisition.

Amber

Instrument is initializing the measurement.

Green

Instrument is collecting data. In Hammer mode, it is looking for a hit.