Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 139

Advertising
background image

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

139

Collecting and Reviewing Data

Chapter 4

Collect Offroute Data using a Pre-defined Measurement

The Enpac 2500 data collector provides seven pre-defined measurements for you
to use to collect Offroute data. These are the data collectors pre-defined
measurements.

You do not set any parameters when you use a pre-defined measurement. The
Enpac 2500 data collector uses similar settings as the measurement definitions for
the current point.

Follow these steps to collect an Offroute measurement for the current point.

1.

From the Main Menu, select Data Collection and press one of the Enter
keys.

The Data Collection screen appears.

2.

Select the route and measurement point.

3.

Press F2 (Offroute) to collect an Offroute measurement.

Table 35 - Pre-defined Measurements

Measurement

Description

Same

Uses the same setup as the current point.

Opposite

Uses the opposite setup as the current point.
For example, spectrum is time, time is spectrum.

Accel

Acceleration spectrum point
10kHZ, 800 lines, 4 averages

Velocity

Accel velocity spectrum point
2kHz, 400 lines, 4 averages

Accel Time

Time point
10kHz, 2046 points

High Res

High resolution point (increase number of lines by a factor of two.
For example, the current point having 400 lines becomes 800 lines.

gSE

Spike energy point
Envelope parameter in Instrument Setup must be set to gSE.
See

Instrument Configuration Parameters on page 44

.

ESP

Standard Envelope Spectrum Processing (ESP) point
Envelope parameter in Instrument Setup must be set to ESP.
See

Instrument Configuration Parameters on page 44

.