4.6 CONT mode - preparation of solutions

711 Liquino, Instructions for use

96

4.6.3

CONT mode parameters

PARAM

B

All settings in the parameter menu form a method and can be stored as
such.

Main menu:

par amet er s

>CONT par amet er

Open submenu with <ENTER>

par amet er s

>CONT dosi ng uni t s

Use <> or <> to move down or up from one menu point to the
next

par amet er s

>CONT r at es

Use <HOME> or <END> to jump to first or last menu point

par amet er s

>CONT pr esel ect i ons

Return to basic condition with <QUIT>

par amet er s

>CONT r epor t s

Submenus:

par amet er s

>CONT par amet er

Dosing parameter settings

Navigate with <ENTER>

>CONT par amet er

t ype: subst . conc

Select the dosing procedure

subst.conc.

mass conc..

mass fraction

volume fraction

= solutions in mol/L
= solutions in g/L
= solutions in %
= dilutions in mol/L, ppm, g/L, %

>CONT par amet er

t ar get cont ent 1. 0 Mol / L

Target concentration

0.001…9999.9

The unit depends on the 'type',

see above.


>CONT par amet er

t ar get vol ume 10. 0 mL

Target volume

0.01…9999

.


>CONT par amet er

f act or 1. 0

Correction factor

0.0000001…999999.0

not for 'volume fraction'.


>CONT par amet er

mol ar mass 1. 0 g/ Mol

Molar mass of the substance

0.0001…999999

only for 'subst.conc'.


>CONT par amet er

densi t y 1. 0 g/ mL

Density of solvent

0.1…99.9999

only for 'mass fraction'.


>CONT par amet er

meas. uni t Mol / L

Mass unit of target content

mol/L, ppm, g/L, %

only for 'volume fraction'.


Advertising