6.2 Error messages

711 Liquino, Instructions for use

144

6.2

Error messages

* 50mL cyl . not possi bl e

In the PIP mode 50 mL dosing units cannot be used. Press <QUIT>

and <EXCHANGE>. Exchange the dosing unit.

* * * bat t er y down

The memory card battery is empty. Insert the card in the instrument
and replace the battery according to the instructions given on the
leaflet accompanying the memory card. Press <QUIT>. After chang-
ing the battery enter the date for the next battery change under
'<MEM CARD>, '>change battery'.

or

* * * bat t er y down

If no memory card is inserted and this message appears when the
Liquino is switched on then the support battery in the 711 Liquino is
empty. Please contact the Metrohm service department.

* * * change bat t er y

The date for the regular memory card battery change has been
reached. Change the battery according to the instructions given on
the leaflet accompanying the memory card. Leave the memory card
inserted in the instrument. Press <QUIT>. After changing the battery
enter the date for the next battery change under '<MEM CARD>,
'>change battery'.

* * * car d f ul l

No more free memory space is available on the memory card. Delete
methods which are no longer needed or use a different card. Press
<QUIT>.

* car d not accessi bl e

The memory card is being accessed simultaneously by the remote
control and by manual operation. Press <QUIT> and try again.

* * * car d r ead/ wr i t e er r or

The Liquino cannot read the memory card or cannot write on it.

Check whether the memory card has been inserted correctly. The
card may have been damaged by external influences. Press
<QUIT>.

* * * car d unf or mat ed

Before a memory card is used for the first time it must be formatted in
the Liquino. Press <QUIT> and format the card under '<MEM
CARD>, >Format'.

* car d wr i t e- pr ot ect ed

The memory card is write-protected. Remove the write protection
(see leaflet accompanying memory card) or use a different card.
Press <QUIT>.

* * * dat a l ost

The inserted memory card is damaged and can no longer be read.
Press <QUIT> and use a different memory card.

* * di r . al r eady exi st s

A directory with the given name already exists. Select another name
for the new directory. Press <QUIT>.

DOSxx: f i l l f i r st !

The piston of Dosino xx is in an undefined position. Press <QUIT>
and then <FILL>.

DOSxx: i nval i d dos. por t

The multiple stopcock of Dosino xx is in an undefined position. Press

<QUIT> and then <FILL> or <EXCHANGE>.

DOSxx: not r eady

Dosino xx is not connected or no dosing unit is attached. The ex-
change unit may not be mounted correctly. Correct the fault and
press <QUIT>.

DOSxx: over l oad!

Dosino xx is blocked. A tubing or the multiple stopcock may be
blocked. Check the tubing and, if necessary, remove the dosing unit
(not the Dosino!). Then press <QUIT>.