3 keyboard and graphics options – Metrohm 711 Liquino User Manual

Page 77

Advertising
background image

4. Detailed description

Liquino 711, Instructions for use

71

>>er r . l i mi t s 20mL- cyl .

Submenu for 20 mL dosing units

er r or max. vol . +/ - 60 µL 1…999 µL

maximum volume error

sl ope +/ - 0. 003 0.001…0.1

maximum variation in slope

i nt er cept +/ - 30 µL 1…99 µL

maximum y-axis intercept

>>er r . l i mi t s 50mL- cyl .

Submenu for 50 mL dosing units

er r or max. vol . +/ - 150 µL 1…999 µL

maximum volume error

sl ope +/ -

0. 003 0.001…0.1

maximum variation in slope

i nt er cept +/ - 50 µL 1…99 µL

maximum y-axis intercept


4.3.3

Keyboard and graphics options

Further configuration possibilities are available:

Locking the keyboard when the Liquino 711 is operated exclusively by

remote control.

Locking individual dialog ranges

Settings for graphics printout

Hiding individual instrument modes


These settings can be made in a special setup menu which is only accessi-
ble when the Liquino is switched on

with the <CONFIG> key pressed

down.



CONFIG

A

Switch the Liquino off, press down the <CONFIG> key and keep this
pressed down while you switch the instrument on again.

Main menu:

set up

>keyboar d opt i ons

open submenu with <ENTER>

set up

>gr aphi cs pr i nt out

use <> or <> to move up or down by 1 menu item

set up

>show/ hi de modes

use <HOME> or <END> to jump to first or last menu item

use <QUIT> to return to the basic condition

Submenus:

set up

>keyboar d opt i ons

Locks keyboard or particular keys

Navigate with <ENTER>

>keyboar d opt i ons

l ock keyboar d: of f

Switches off keyboard

on

off

= the Liquino cam only be operated via external software.


If the keyboard is locked the setup menu is still available to switch off the
keyboard lock.

Advertising