3.2 Tutorial

711 Liquino, Instructions for use

36



<ENTER>







<ENTER>





<ENTER>

di l ut i on vol . 3 mL

The dilution volume is the amount of solvent to be added to the meas-

uring cell after the sample has been transferred there (in the example 3
mL). Press <ENTER> to accept the value.

Di spl ay:

>SMPL sampl i ng oper at i on

sampl e r i nse vol . 3 mL

Before the actual sample transfer the Dosino cylinder is rinsed with the

sample. This rinsing volume should be about the same as the transfer
amount (in the example 3 mL). For a multiple determination, in which
the same sample is always used, this volume can also be smaller.
Press <ENTER> to accept the value.

Di spl ay:

>SMPL sampl i ng oper at i on

sampl e r i nse r ep. 1

Enter how often the cylinder is to be rinsed with the sample. If samples

are being transferred which do not greatly differ from one another 1
rinse is sufficient. Press <ENTER> to accept the value.

QUIT

NO

As the remaining parameters in this submenu can be left at their preset

values it can be exited with the <QUIT> key.

Di spl ay:

par amet er s

>SMPL sampl i ng oper at i on

J


<ENTER>





<ENTER>





<ENTER>







<ENTER>






<ENTER>

The next submenu is accessed with <>.

Di spl ay:

par amet er s

>SMPL meas. cel l r i nsi ng

Press <ENTER> to open the submenu ‘SMPL meas.cell rinsing.

Di spl ay:

>SMPL meas. cel l r i nsi ng

ext er nal Pump: of f

The cell can also be rinsed with external pumps. However, the same

Dosino is to be used here. Press <ENTER> to accept the default val-
ue ‘off’.

Di spl ay:

>SMPL meas. cel l r i nsi ng

r i nsi ng r epet i t i ons 1

Enter the number of times that the measuring cell is to be rinsed. As

this process takes a considerable time this value should be kept as low
as possible. In this example 1 rinse is sufficient. Press <ENTER> to
accept the value.

Di spl ay:

>SMPL meas. cel l r i nsi ng

aspi r . vol ume 40 mL

The suction volume corresponds to the sample volume + dilution vol-

ume (solvent) + titrant consumption. The value must be so large that
the measuring cell is completely emptied after the titration. In this case
the total volume amounts to 40 mL. Press <ENTER> to accept the
value.

Di spl ay:

>SMPL meas. cel l r i nsi ng

aspi r . r at e max mL/ mi n

This setting is left at ‘max’. It only requires alteration for viscous liquids.

Press <ENTER>.