3.2 Tutorial

711 Liquino, Instructions for use

56

Running the method

START

Z

When <START> is now pressed the display shows

Di spl ay:

Not pr epar ed yet !

Run PREP now?

ENTER

YES

Before the Dosino is used for the first time in the new configuration the

Dosino cylinder and the complete tubing system must be filled bubble-
free with the particular reagent solution. Press <YES>.

Di spl ay:

wast e por t :

t i p( 1) or f l ask( 2) ?

Press <1> to empty the contents of the cylinder through the dosing tip

or <2> to return them to the bottle. Take care that the tips of the dos-
ing tubing is located in a vessel. Wait until the 711 Liquino has ended
the preparation (PREP) of all the Dosinos which are being used. The
instrument then returns to its normal state.

START

Z

From now on when <START> is pressed the method will be run im-

mediately.

*)



*) The warning about the preparation of the tubing system appears each time the mode is changed and after
the instrument has been switched off and on again. If it is certain that the Dosinos used have already been
prepared then press <NO> (<QUIT> key) to acknowledge the warning. The method can now be run in the
normal manner by pressing <START>.