8 check remote interfaces, 9 check rs232 interfaces – Metrohm 711 Liquino User Manual

Page 160

Advertising
background image

6.5 Diagnosis

711 Liquino, Instructions for use

154

6.5.8

Check remote interfaces

This diagnosis step allows the functioning of all outputs and inputs as well

as the input filter of the remote interface to be checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>r emot e t est

<ENTER>

>r emot e t est

>I O CONNECTOR ?

Without switching the instrument off insert test plug 3.496.8510 into

the "Remote" socket.

<ENTER>

The test proceeds automatically. If no faults are found the display

shows:

>r emot e t est

I O TEST OK

Remove test plug.

The test is exited by pressing the <STOP> key.

di agnosi s

>RS232 t est

6.5.9

Check RS232 interfaces

This diagnosis step allows the functioning of all outputs and inputs of the

RS232 interface to be checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>RS232 t est

<ENTER>

>RS232 t est

RS CONNECTOR ?

Without switching the instrument off insert test plug 3.496.8560 into

the "RS232" socket.

<ENTER>

The test proceeds automatically. If no faults are found the display

shows:

>RS232 t est

RS TEST OK

Remove test plug.

The test is exited by pressing the <STOP> key.

di agnosi s

>ext er nal bus t est