2.5.4 Connecting a balance


The following balances can be connected to the RS232 output of the
711 Liquino:

Balance

Cable

Sartorius MP8, MC1

6.2134.060

Mettler AB, AG (LC-RS25)

supplied with balance

Mettler AM, PM

from Mettler: ME 33995: green wire to Pin 2, brown wire to
Pin 3, white wire to Pin 7, yellow wire to Pin 20 of the 25-pin
plug
+ 6.2125.010 25-pin/9-pin adapter

Mettler interface 016

cable supplied with interface 016: red wire to Pin 3, white
wire to Pin 7 of the 25-pin plug
+ 6.2125.010 25-pin/9-pin adapter

Mettler interface 011 or 012

6.2125.020 + 6.2125.010

Mettler AT

from Mettler: ME 33995: green wire to Pin 2, brown wire to
Pin 3, white wire to Pin 7, yellow wire to Pin 20 of the 25-pin
plug
+ 6.2125.010 25-pin/9-pin adapter

AND models ER-60, 120, 180, 182

models FR-200, 300

models FX-200, 300, 320

with RS232 interface (OP-03)

6.2125.020 + 6.2125.010

Precisa, balances with RS232C inter-
face

6.2125.080 + 6.2125.010


The balance type must be preselected on the 711 Liquino with the
<CONFIG> key under ‘>Peripheral units’.
The weight will be transmitted as a number with up to 6 places, together
with sign and decimal point. Units and control characters sent by the bal-
ance will not be transmitted.

For the simultaneous connection of printer and balance the adapter cable
6.2125.010 + with junction plug 6.2125.030 are required.
If the balance only works with 7 bit and the printer only works with 8 bit then
the balance parity must be set to "space" while the 711 Liquino and printer
run with 8 bit, "no parity".