4. Detailed description

711 Liquino, Instructions for use

111

Rinse Dosino

rinse

Dosino

repeat

rinsing

empty

Dosino



The Dosino aspirates the sample solution. The aspi-

rated amount is defined by the sample rinsing volume
and the tubing volume at port 1.

The dosing cylinder is emptied into the waste bottle

via port 4.




This initial rinsing can be repeated up to 10 times.





Add auxiliary solution

add aux.

solutions


With further Dosinos up to three different auxiliary

solutions can be added, see '>SMPL Dosing units'.
After the addition of each auxiliary solution a waiting
period can be programmed (in HH:MM:SS). The ad-
dition of auxiliary solutions is optional. It is possible
before or after sample transfer.

Transfer sample

aspirate

sample

add

sample

The Dosino aspirates the sample via port 1.



The defined sample volume is dosed into the measur-

ing cell (via port 2).

Dilute sample

aspirate

solvent

add

solvent

The Dosino aspirates the solvent via port 3.




The defined dilution volume is dosed into the measur-

ing cell (port 2).


Advertising