Metrohm 711 Liquino User Manual

Page 31

Advertising
background image

3 Introduction

711 Liquino, Instructions for use

25

If you do not want the dosing speed to increase or you require a quicker

increase in the dosing speed then you can set this in the configuration.

CONFIG

A

2x

J













<ENTER>






<ENTER>







<SELECT>

<ENTER>






<ENTER>

Press the <CONFIG> key and the <> key twice.

Di spl ay:

conf i gur at i on

>manual

Here you will find the submenu with the settings which are effective for

manual functions.

Di spl ay:

>manual DOS1

← dosing connection

dos. uni t : def aul t

In the first line, above right, the address of the dosing connection to

which the particular setting applies is shown. You can make different
settings for each of the four dosing connections. The settings for the
other dosing connections DOS2 to DOS4 are also found in this sub-
menu. Now press the <ENTER> key three times.

Di spl ay:

>manual DOS1

>>dosi ng

If you want to alter the dosing behavior of dosing device 1 you will find

this in the second submenu '>>Dosing' after pressing the <ENTER>
key as given above. If your Dosino is connected to 'Dos. 2' then press
the <ENTER> key six times.

Di spl ay:

>>dosi ng DOS1

r at e t ype: i ncr easi ng


By pressing the <SELECT> key and confirming with <ENTER> you

can switch to a constant dosing speed. You should, however, leave the
setting with increasing speed alone and press <ENTER>.

Di spl ay:

>>dosi ng DOS1

sl ope 4

Enter the highest value '10' for the slope of the increase of the dosing

speed. Confirm the entry with <ENTER>.

By pressing the <QUIT> key twice you will return to the display for dos-

ing.

Di spl ay:

manual DOSI NG

r eady 3. 349 mL

START

Z

Start dosing again by pressing the <START> key; you will see that the

dosing speed increases more rapidly and that the highest dosing speed is
reached very quickly.

FILL

5

M

When the max. cylinder volume has been reached the dosing unit will be

filled automatically. Filling the dosing cylinder can also be triggered manu-
ally by pressing the <FILL> key.

Advertising