Metrohm 711 Liquino User Manual

Page 45

Advertising
background image

3 Introduction

711 Liquino, Instructions for use

39

<ENTER>

<QUIT>

Di spl ay:

par amet er s

>SMPL pr esel ect i ons

As the ‘SMPL control I/O’ submenu contains no further menu items, an

automatic return is made to the submenu selection. The parameter
menu is exited with <QUIT>. The instrument is now ready to carry out
the SMPL method.



Various SMPL methods can be edited and then stored in the method
memory. Each of these methods can then be reloaded by pressing a few
keys. The method memory is described on page 128.



Running the method

START

Z

When <START> is now pressed the display shows

Di spl ay:

Not pr epar d yet !

Run PREP now?

ENTER

YES

Before the Dosino is used for the first time in the new configuration the

Dosino cylinder and the complete tubing system must be filled bubble-
free with the particular reagent solution. Press <YES>.

Di spl ay:

wast e por t :

t i p( 1) or f l ask( 2) ?

Press <1> to empty the contents of the cylinder through the dosing tip

or <2> to return them to the bottle. Take care that the tip of the dosing
tubing is located in a vessel. Wait until the 711 Liquino has ended the
preparation (PREP) of all the Dosinos which are being used. The in-
strument then returns to its normal state.

START

Z

From now on when <START> is pressed the method will be run im-

mediately.

*)

*) The warning about the preparation of the tubing system appears each time the mode is changed and after
the instrument has been switched off and on again. If it is certain that the Dosinos used have already been
prepared then press <NO> (<QUIT> key) to acknowledge the warning. The method can now be run in the
normal manner by pressing <START>.