4. Detailed description

711 Liquino, Instructions for use

115

>SMPL meas. cel l r i nsi ng

r i nsi ng vol ume 50. 0 mL

Solvent volume for rinsing the cell

0.1…250

with 'external pump off'


>SMPL meas. cel l r i nsi ng

r i ns. r at e max mL/ mi n

Dosing rate for rinsing the cell

0.001…166.66,

max

with 'external pump off'


= highest possible dosing rate (depends on type of
dosing unit), select with <CLEAR> key


par amet er s

>SMPL dosi ng uni t s

Settings for dosing units and auxiliary dosing devices

Navigate with <ENTER>

>SMPL dosi ng uni t s

dos. uni t 1: def aul t

Name of dosing unit used

selection list

see configuration '

>dosi ng uni t s'

>SMPL dosi ng uni t s

>>aux. sol ut . BEFORE

Submenu for definition of dosing proce-

dures before sample transfer

>>aux. sol ut . BEFORE no1

dosi ng dr i ve: of f

Dosing connection for addition of first
auxiliary solution

DOS2, DOS3, DOS4, off

>>aux. sol ut . BEFORE no1

dos. uni t : def aul t

Name of dosing unit used

selection list

see configuration '

>dosi ng uni t s'

>>aux. sol ut . BEFORE no1

vol ume 1. 000 mL

Volume of first auxiliary solution to be
added

0.0001…99999.9

>>aux. sol ut . BEFORE no1

f eed r at e max mL/ mi n

Dosing rate of auxiliary dosing device

0.001…166.66,

max


= highest possible dosing rate (depends on type of
dosing unit), select with <CLEAR> key

>>aux. sol ut . BEFORE no1

f i l l r at e max mL/ mi n

Filling rate of auxiliary dosing device

0.001…166.66,

max


= highest possible filling rate (depends on type of
dosing unit), select with <CLEAR> key

>>aux. sol ut . BEFORE no1

wai t t i me 00: 00: 00

Waiting time after addition of auxiliary
solution

HH:MM:SS

>>aux. sol ut . BEFORE no2

dosi ng dr i ve: of f

Dosing connection for addition of second
auxiliary solution

DOS2, DOS3, DOS4, off

…up to 3 auxiliary dosing

devices