Metrohm 711 Liquino User Manual

Page 58

Advertising
background image

3.2 Tutorial

711 Liquino, Instructions for use

52

For the other two XDOS methods (diazotization and azo coupling) the pa-
rameter lists are printed out below. Use these lists together with the practical
experience already gained to enter the parameters. Lines marked with

( adapt ) must be adapted to the existing conditions.

XDOS method for diazotization

711 Li qui no 711. 0011

dat e 1998- 07- 08 t i me 18: 16

mode: XDOS user met hod: NaNO2add

par amet er s

>XDOS par amet er

XDOS t ype: vol ume&r at e

vol ume 80. 0 mL

f eed r at e 30. 0 mL/ mi n

t emp. measur i ng on

r ec. i nt er val 00: 01: 00

f i l l r at e max mL/ mi n

aut o f i l l on

>XDOS dosi ng uni t s

set up: si ngl e

dosi ng dr i ve: DOS2

( adapt )

dos. uni t : NaNO2

( adapt )

por t use: def aul t

>XDOS moni t or i ng

t emper at ur e: on

l ow l i mi t - 70 ° C

upper l i mi t 4. 0 ° C

act i on: wai t

al ar m pi ns: none

vol ume: of f

>XDOS pr esel ect i ons

r eq. i dent : of f

>>XDOS pr ompt s

i d1 I d1

i d2 I d2

i d3 I d3

st ar t del ay 00: 00: 00

>XDOS r epor t s

t emp. aut o scal e: on

r epor t 1: r esul t

( wi t h pr i nt er , i f not : of f )

r epor t 2: cur ve

( wi t h pr i nt er , i f not : of f )

r epor t 3: of f

r epor t 4: of f

r epor t 5: of f

r epor t 6: of f

r epor t 7: of f

r epor t 8: of f

============