9 xdos – the electronic dropping funnel – Metrohm 711 Liquino User Manual

Page 49

Advertising
background image

3 Introduction

711 Liquino, Instructions for use

43

3.2.9

XDOS – the electronic dropping funnel


120mL of solution are to be added in 5 hours. This can be carried out easily
and reliably with a tandem addition using 2 Dosinos. In this way interruption-
free addition is guaranteed.

Preconditions / Preparation

The accessories for tandem addition (see ”ordering information” in the ”711
Liquino” leaflet) are required for the tubing. However, it is also possible to
lead a separate tubing from each Dosino to the test apparatus.

A dosing unit which corresponds to the tubing used for the Dosino must be
defined. This process is described on page 22.

Method editing

*

MODE

-

X

DOS NR

SELECT

P

The XDOS mode must first be activated. This sets all method parame-

ters to standard values. Press the <MODE> key and then <SELECT>
until ‘XDOS’ appears in the display:

Di spl ay:

mode sel ect : XDOS

ENTER

YES

Press <ENTER>. The instrument is now in the XDOS mode.

Di spl ay:

XDOS * * * * * * * *

r eady 0. 000 mL

PARAM

B

The <PARAM> key is used to access the parameter menu, where all

the possible settings for an XDOS method can be found.

Di spl ay:

par amet er s

>XDOS par amet er s

The submenu ‘XDOS parameters’ contains the main settings for an

XDOS method. Press <ENTER> to open the submenu.

Di spl ay:

>XDOS par amet er s

XDOS t ype: vol ume&r at e

The colon indicates that the <SELECT> key can be used to select a

value from a list. Press <SELECT> until ‘volume&time’ appears. Ac-
cept the entry with <ENTER>.

Di spl ay:

>XDOS par amet er s

vol ume 1. 000 mL

Enter the volume which is to be added (1000 mL in this example) and

press <ENTER>.

Di spl ay:

>XDOS par amet er s

dosi ng t i me 00: 01: 00

Enter the length of time for the addition in ‘hours:minutes:seconds’

format (if a mistake is made press the <CLEAR> key and repeat the
entry). Accept the value with <ENTER>.

Di spl ay:

>XDOS par amet er s

t emp. measur i ng: of f

QUIT

NO

No further entries are necessary in the ‘XDOS parameters’ submenu.

Exit with the <QUIT> key.

Di spl ay:

par amet er s

Advertising