4. Detailed description

711 Liquino, Instructions for use

79

par amet er s

>XDOS moni t or i ng

Monitoring function settings

Navigate with <ENTER>

>XDOS moni t or i ng

t emper at ur e: of f

Temperature monitoring

on, off

>XDOS moni t or i ng

l ow l i mi t - 70 ° C

Lower temperature limit

-70…200.0

>XDOS moni t or i ng

upper l i mi t 200 ° C

Upper temperature limit

-70…200.0

>XDOS moni t or i ng

act i on: none

Reaction if upper or lower temperature limit is
infringed

none,

stop,

hold

wait

= no reaction (alarm pins still active)
= dosing is stopped
= dosing is interrupted and can be continued with
<CONT>
= dosing is interrupted and automatically continued
as soon as the measuring temperature is again within
the limits.

>XDOS moni t or i ng

al ar m pi ns: none

Definition of the active alarm pins of the re-
mote connection

none,

under,

over

all

= no activation
= Output2 lead (Lower Limit, pin 4) activated when tem-
perature too low
= Output1 lead (Upper Limit, pin 18) activated when
temperature too high
= Output1 and Output2 activated when corresponding
temperature limits are infringed (see above)

>XDOS moni t or i ng

vol ume: of f

Volume monitoring function

on, off

>XDOS moni t or i ng

cr i t . vol ume 10000 mL

Absolute volume limit for terminating dosing

0.0…99999.9

onl y f or ' vol ume: on'


par amet er s

>XDOS pr esel ect i ons

Definition of requests at start of dosing

Navigate with <ENTER>

>XDOS pr esel ect i ons

r eq. i dent : of f

Automatic request of up to
3 identification texts

id1,

id1&2,

all,

off

= only ID1 is requested
= ID1 and ID2 are requested
= ID1, ID2 and ID3 are requested
= no request

Use identification texts (ID1…3) for identifying your sample or your test. The
IDs will be printed out in the report of the dosing procedure so that the indi-
vidual result reports or curves are identified unambiguously.

The alarm pins can be

used to switch an external

instrument connected to

the remote socket, e.g. a

heating bath or a

thermostat.

You are recommended to

use Metrohm Relay Box

731.