6. Appendix

711 Liquino, Instructions for use

153

6.5.6

Check display contrast

This diagnosis step allows the contrast setting variability function to be

checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>di spl ay cont r ast t est

<ENTER>

After pressing the <ENTER> key the following display appears con-

tinuously with changing contrast settings for an optical check of the

display.

>di spl ay cont r ast t est

The test is exited by pressing the <STOP> key.

di agnosi s

>key t est

6.5.7

Check keyboard

This diagnosis step allows the functioning of all the keyboard keys to be

checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>key t est

<ENTER>

>key t est

Press all keys in sequence and observe the reaction on the display

with the help of the following keyboard matrix.

Fig. 1

Keyboard matrix

The corresponding matrix code for the pressed-down key appears in the

display. The matrix code can be seen from the keyboard matrix along-

side.

For example, when the <CONFIG> key is pressed the following display

must appear:

>key t est

KEY: 31 CONFI G

The test is exited by pressing the <STOP> key twice.

di agnosi s

>r emot e t est