4.9 GLP mode - validation

711 Liquino, Instructions for use

122

ceptance can be previously set in the parameter menu under '>GLP Op-
tions'.
Ten different and randomly chosen volumes will be dosed out. Enter the
weight each time. At the end of the GLP mode the defined GLP report will
be printed out.

In the result or detailed report the Y-axis intercept and the slope of the linear
regression (theoretical volume against actual volume) and the volume error
(referred to the complete cylinder volume) will be indicated.
The determined values are compared with the set limiting values by the
Liquino and a corresponding commentary is produced; see example of re-
port on page 125.