Metrohm 711 Liquino User Manual
Page 54
3.2 Tutorial
711 Liquino, Instructions for use
48
<SELECT>
<ENTER>
<ENTER>
<ENTER>
<SELECT>
<ENTER>
<ENTER>
<SELECT>
<ENTER>
<QUIT>
2x <>
<ENTER>
dos. uni t : st andar d
•
In this menu item the 711 Liquino is informed about the tubing of the
Dosino. Press <SELECT> until the dosing unit defined in the above
preparation step appears. Confirm with <ENTER>.
•
Skip the next menu item with <ENTER>.
Di spl ay:
par amet er s
>XDOS moni t or i ng
•
As the submenu ‘>XDOS dosing units' does not contain any further
menu items, an automatic return is made to the submenu selection.
Press <ENTER> to open the submenu ‘>XDOS monitoring'.
Di spl ay:
>XDOS moni t or i ng
t emper at ur e: of f
•
Use the <SELECT> key to select the entry ‘on’. This activates the
temperature monitoring.
Di spl ay:
>XDOS moni t or i ng
l ow l i mi t - 70 ° C
•
The lower temperature limit is of no interest in this application. Accept
the default value with <ENTER>.
Di spl ay:
>XDOS moni t or i ng
upper l i mi t 200 ° C
•
Set the upper temperature limit to 10 °C. Confirm the entry with
<ENTER>.
Di spl ay:
>XDOS moni t or i ng
act i on: none
•
The action to be taken if the temperature exceeds the set limits is de-
termined here. Use <SELECT> to choose the entry ‘wait’. This stops
any addition when the temperature is outside the limits. When the tem-
perature again lies within the limits the addition will be continued. Ac-
cept the setting with <ENTER>.
•
No further entries are necessary in the ‘XDOS monitoring’ submenu.
Exit with the <QUIT> key.
Di spl ay:
par amet er s
XDOS moni t or i ng
•
Press <> 2x to switch to the submenu ‘XDOS reports’.
Di spl ay:
par amet er s
>XDOS r epor t s
•
Press <ENTER> to open the submenu.
Di spl ay:
>XDOS r epor t s
t emp. aut o scal e: on
*) If no printer or PC is connected the parameter menu can be exited with <QUIT> and the method rejoined
at "Running the method“.