3.2 Tutorial

711 Liquino, Instructions for use

48

<SELECT>

<ENTER>

<ENTER>

<ENTER>

<SELECT>

<ENTER>

<ENTER>

<SELECT>

<ENTER>

<QUIT>

2x <>

<ENTER>

dos. uni t : st andar d

In this menu item the 711 Liquino is informed about the tubing of the

Dosino. Press <SELECT> until the dosing unit defined in the above
preparation step appears. Confirm with <ENTER>.

Skip the next menu item with <ENTER>.

Di spl ay:

par amet er s

>XDOS moni t or i ng

As the submenu ‘>XDOS dosing units' does not contain any further

menu items, an automatic return is made to the submenu selection.
Press <ENTER> to open the submenu ‘>XDOS monitoring'.

Di spl ay:

>XDOS moni t or i ng

t emper at ur e: of f

Use the <SELECT> key to select the entry ‘on’. This activates the

temperature monitoring.

Di spl ay:

>XDOS moni t or i ng

l ow l i mi t - 70 ° C

The lower temperature limit is of no interest in this application. Accept

the default value with <ENTER>.

Di spl ay:

>XDOS moni t or i ng

upper l i mi t 200 ° C

Set the upper temperature limit to 10 °C. Confirm the entry with

<ENTER>.

Di spl ay:

>XDOS moni t or i ng

act i on: none

The action to be taken if the temperature exceeds the set limits is de-

termined here. Use <SELECT> to choose the entry ‘wait’. This stops
any addition when the temperature is outside the limits. When the tem-
perature again lies within the limits the addition will be continued. Ac-
cept the setting with <ENTER>.

No further entries are necessary in the ‘XDOS monitoring’ submenu.

Exit with the <QUIT> key.

Di spl ay:

par amet er s

XDOS moni t or i ng

Press <> 2x to switch to the submenu ‘XDOS reports’.

*)

Di spl ay:

par amet er s

>XDOS r epor t s

Press <ENTER> to open the submenu.



Di spl ay:

>XDOS r epor t s

t emp. aut o scal e: on

*) If no printer or PC is connected the parameter menu can be exited with <QUIT> and the method rejoined
at "Running the method“.