15 check temperature calibration – Metrohm 711 Liquino User Manual

Page 164

Advertising
background image

6.5 Diagnosis

711 Liquino, Instructions for use

158

Press <ENTER>.

After pressing the <ENTER> key the interface Dos.1 will be ad-

dressed and information about the connected Dosino shown on the

display.

>dosi no t est

Dos: 1, I NFO: 0x??, CYL: ??

If no dosing device is connected

Dos:1, INFO:0x03, CYL: 0

must appear in the display.

Check the information in the second line against Table 1 or Table 2.

Press <ENTER>.

After pressing the <ENTER> key the interface Dos.2 will be ad-

dressed and information about the connected Dosino shown on the

display.

In the same way as for Dos.1 the interfaces Dos.2, Dos.3 and Dos.4

can now be checked in sequence.

Dosing device

Information

700 Dosino with

I NFO: 0xBB, CYL: 0

700 Dosino with 2 mL

I NFO: 0xB3, CYL: 2

700 Dosino with 5 mL

I NFO: 0xAB, CYL: 5

700 Dosino with 10 mL

I NFO: 0xA3, CYL: 10

700 Dosino with 20 mL

I NFO: 0x9B, CYL: 20

700 Dosino with 50 mL

I NFO: 0x93, CYL: 50

Code if a 700 Dosino is connected

The test is exited by pressing the <STOP> key.

di agnosi s

>t emper at ur e cal i br at i on

6.5.15

Check temperature calibration

This diagnosis step allows the functioning of the interface for recording the

temperature to be checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>t emper at ur e cal i br at i on

<ENTER>

After pressing the <ENTER> key the temperature sensor will be ad-

dressed and shown on the display.

>t emper at ur e cal i br at i on

OVERRANGE, CHECK SENSOR

Insert temperature sensor into "Pt100" socket.

>t emper at ur e cal i br at i on

TEMP. CELSI US ??. ?

Check whether the displayed value coincides with the actual temper-

ature.

The test is exited by pressing the <STOP> key.

di agnosi s

>power on r eset