5.2 Please note …

711 Liquino, Instructions for use

136

chloromethane, trinitrophenol-2,3,6, vinyl acetate, vinyl chloride, vinyli-
dene chloride

Never dose

hot (>60°C) chemicals or solutions with Dosinos!

Metrohm offers tubing made of

FEP (tetrafluoroethylene-

hexafluoropropylene copolymer) and

PTFE (polytetrafluoroethylene).

These materials are largely resistant to organic solvents, strong acids
and alkalis.
FEP tubing is not suitable for:

perchloric acid

fluorine

FEP tubing has a limited resistance to:

chlorosulfonic acid

FEP tubing has a limited resistance at higher temperatures to:

crude oil products

perchloroethylene

tetrachloroethylene

PTFE tubing has a limited resistance to:

halogenated hydrocarbons

perchloric acid

fluorine

PTFE tubing has a limited resistance at higher temperatures to:

crude oil products

perchloroethylene

tetrachloroethylene


Unoccupied dosing unit connections (Ports) can be sealed with

drying tubes or stoppers.

Divide your synthesis run into part steps and create separate XDOS

methods for the individual dosing steps. Then you can link together the
individual part steps (methods) to form a sequence (SEQ mode) and
add further commands to it.