1 xdos mode parameters – Metrohm 711 Liquino User Manual

Page 83

Advertising
background image

4. Detailed description

711 Liquino, Instructions for use

77

4.4.1

XDOS mode parameters

PARAM

B

All settings in the parameter menu form a method and can be stored as
such.

Main menu:

par amet er s

>XDOS par amet er

Open submenu with <ENTER>

par amet er s

>XDOS dosi ng uni t s

Use <> or <> to move down or up from one menu point to the
next

par amet er s

>XDOS moni t or i ng

Use <HOME> or <END> to jump to first or last menu point

par amet er s

>XDOS pr esel ect i ons

Return to basic condition with <QUIT>

par amet er s

>XDOS r epor t s


Submenus:

par amet er s

>XDOS par amet er

Dosing parameter settings.

Navigate with <ENTER>

>XDOS par amet er

XDOS t ype: vol -

ume&r at e

Selection of task parameters

volume&rate,

volume&time,

time&rate

= selection of volume and dosing rate
= selection of volume and dosing time
= selection of dosing time and dosing rate

>XDOS par amet er

vol ume 1. 0 mL

Dosing volume

0.0…99999.9,

infinite

not for time&rate


= continuous dosing; can be ended with <STOP>

>XDOS par amet er

f eed r at e max mL/ mi n

Dosing rate

0.001…166.66,

max

not for volume&time


= highest possible dosing rate (depends on type of

dosing unit.)

The dosing rate is the adjustable piston stroke speed when discharging the
cylinder contents. For simple dosing it does not correspond to the effective
delivery rate as the filling times of the dosing unit must be taken into ac-
count.

>XDOS par amet er

dosi ng t i me 00: 01: 00

Time selected for complete dosing process.
Filing times must be taken into account.

HH:MM:SS

not for volume&rate

in hours, minutes and seconds (corrections can be made
with <CLEAR>.)

>XDOS par amet er

t emp. measur i ng: of f

Temperature measurement (compulsory for
temperature-controlled dosing)

on, off


>XDOS par amet er

r ec. i nt er val 00: 01: 00

Recording interval for
temperature measurement

HH:MM:SS

(only for temp. measuring 'on'

)

in hours, minutes and seconds

Depending on the

selected XDOS type

two of three dosing

parameters will

be shown.

The third parameter will

be automatically

determined by the

Liquino 711 or results

from the dosing itself

and will be printed out in

the result report when

dosing has been con-

cluded.