4 installing the pipetting attachment, 4 memory card – Metrohm 711 Liquino User Manual

Page 14

Advertising
background image

2.4 Memory card

711 Liquino, Instructions for use

8

2.3.4

Installing the pipetting attachment


The pipetting attachment available as an option is suitable for automatic pi-
petting functions with a Dosino 700.



To install the pipetting attachment proceed as follows:

Loosen the screw nipple on the pipet handle and push it over the pipet

tubing.

Guide the pipet tubing into the handle and fix the tubing with the screw

nipple.

Mount the pipet tubing on dosing port 1 of the Dosino dosing unit.

Insert in the 25-pin connection cable plug into the remote socket of the

Liquino. The second connection cable plug is used to connect another
peripheral unit to the Liquino.

As soon as the pipet handle is connected the green pilot lamp lights up

to indicate that it is ready for work.

The angle between the tubing guide and the handle can be adjusted for

strain-free work.

Important: the pipet tubing must always be clean and kink-free!

Pipetting accessories:
Pipet handle

6.1562.040

with spiral pipet tubing, 3 mL

6.1562.050

Pipet tubing, 10 mL

6.1562.100

Pipet tubing, 3 mL

6.1829.020

Pipet tubing, 0.7 mL

6.1562.010

Pipet tips for pipet tubing 0.7 mL, 10 pieces

6.1562.030

Pipet tip holder

6.2052.010

2.4

Memory card

The memory card (e.g. 6.2245.010) must be inserted in the slot on the front
panel of the 711 Liquino so that the front of the memory card (with the type
name) faces to the right. A little pressure is necessary to push the memory
card completely into the slot. Never remove the memory card from the in-
strument when it is being accessed by the instrument.

Pipet tip

Green pilot lamp

6.1562.040

Pipet handle

Black start
button

6.1562.050
Pipetting tubing

Connection to dosing port 1
of the Dosino 700

Connection cable to 711 Liquino (remote
connection)

Screw nipple