Metrohm 711 Liquino User Manual

Page 60

Advertising
background image

3.2 Tutorial

711 Liquino, Instructions for use

54

SEQ method editing


As the sub-methods have now been defined and stored, the SEQ method
can now be assembled.

*

MODE

-

X

DOS NR

SELECT

P

The SEQ mode must first be activated. This sets all method parameters

to standard values. Press the <MODE> key and then <SELECT> until
‘SEQ’ appears in the display:

Di spl ay:

mode sel ect : SEQ

ENTER

YES

Press <ENTER>. The instrument is now in the SEQ mode.

Di spl ay:

SEQ * * * * * * * *

r eady 0. 000 mL

PARAM

B

<ENTER>

<SELECT>

<ENTER>


<SELECT>

<ENTER>

<SELECT>

<ENTER>

The <PARAM> key is used to access the parameter menu, where all

the possible settings for a SEQ method can be found.

Di spl ay:

par amet er s

>SEQ sequence

In the submenu ‘SEQ sequence’ previously defined and stored XDOS

methods can be assembled to form a sequence and the correspond-
ing pauses inserted. Press <ENTER> to open the submenu.

Di spl ay:

>SEQ sequence st ep 1

command: NO OPERATI ON

"1. Temperature-controlled acidification of aniline with semi-

concentrated hydrochloric acid (temperature below approx. 10 °C)“.

Use the <SELECT> key to choose the command ‘METHOD’ and con-
firm with <ENTER>.

Di spl ay:

>SEQ sequence st ep 1

met hod:

Press the <SELECT> key until the name of the method which has
been stored for the acidification appears. Press <ENTER> to accept
the selection.

Di spl ay:

>SEQ sequence st ep 2

command: NO OPERATI ON

"2. Pause for 5 minutes for cooling down to 0 °C.“

The <SELECT> key can now be used to select the command ‘PAUSE’
and confirm it with <ENTER>.

Di spl ay:

>SEQ sequence st ep 2

pause 00: 00: 00

Enter a pause of ’00:05:00’. (If a mistake is made press the <CLEAR>
key to delete the value and repeat the entry.)

<SELECT>

<ENTER>

Di spl ay:

>SEQ sequence st ep 3

command: NO OPERATI ON

"3. Temperature-controlled diazotization (temperature below 5 °C).“

Select the command ‘METHOD’ as in step 1 and then the XDOS meth-
od which has been stored for diazotization. Confirm each with
<ENTER>.

Di spl ay:

>SEQ sequence st ep 4

Advertising