10 check external bus, 11 check beeper, 12 check memory card – Metrohm 711 Liquino User Manual

Page 161

Advertising
background image

6. Appendix

711 Liquino, Instructions for use

155

6.5.10

Check external bus

This diagnosis step allows the functioning of those parts of the hardware

for controlling the Dosinos which are contained in the 711 Liquino to be
checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>ext er nal bus t est

<ENTER>

When the <ENTER> key is pressed the Dosino interfaces located on

the internal bus are searched for. During the test the following dis-

play appears:

>ext er nal bus t est

EBUS TESTI NG

The test proceeds automatically. If no faults are found the display

shows:

>ext er nal bus t est

ADDRESS 0x81 TYPE 3

<ENTER>

>ext er nal bus t est

ADDRESS 0x82 TYPE 3

The test is exited by pressing the <STOP> key.

di agnosi s

>beeper t est

6.5.11

Check beeper

This diagnosis step allows the functioning of the internal beeper to be

checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>beeper t est

<ENTER>

When the <ENTER> is pressed the internal beeper is continually

switched on and off. The following display appears during the test:

>beeper t est

The test is exited by pressing the <STOP> key.

di agnosi s

>memor y car d t est

6.5.12

Check memory card

This diagnosis step allows both the functioning of the memory card inter-

face as well as the memory card itself to be checked. The test duration
depends on the size of the memory card to be tested. The test takes 2
minutes for a 128 Kbyte SRAM card.