4. Detailed description

711 Liquino, Instructions for use

123

4.9.1

GLP mode parameters

PARAM

B

All settings in the parameter menu form a method and can be stored as
such.

Main menu:

par amet er s

>GLP par amet er s

Open submenu with <ENTER>

par amet er s

>GLP opt i ons

Use <> or <> to move down or up from one menu point to the
next

par amet er s

>GLP r epor t s

Use <HOME> or <END> to jump to the first or last menu point

Return to normal state with <QUIT>

Submenus:

par amet er s

>GLP par amet er s

Dosing parameter settings

Navigate with <ENTER>

>GLP par amet er s

dosi ng dr i ve: DOS1

Dosing connection

DOS1, DOS2, DOS3, DOS4

>GLP par amet er s

dos. uni t : def aul t

Name of dosing unit used

selection list see configuration '

>dosi ng uni t s'

>GLP par amet er s

t emper at ur e 20. 00° C

Temperature of liquid

-70…200.0

par amet er s

>GLP opt i ons

Other settings

Navigate with <ENTER>

>GLP opt i ons

l i qui d: wat er

Type of liquid

water, other

>GLP opt i ons

l i qu. name

Name of liquid

13 ASCII characters

(only for 'liquid: other')

Text input with <ALPHA> key

>GLP opt i ons

l i qu. dens. 1. 0000 g/ mL

Density of liquid at the measuring tem-
perature

0.0100…20.0000

(only for 'liquid: other')


>GLP opt i ons

f eed r at e max mL/ mi n

Dosing rate

0.001…166.66,

max

(only for 'liquid: other')


= highest possible dosing rate (depends on type
of dosing unit.)