4.8 SMPL mode - fully automatic sampling

711 Liquino, Instructions for use

114

>SMPL sampl i ng oper at i on

sampl e r i nse r ep. 3

Number of times for pre-rinsing dosing
cylinder

1…10 recommended: 1 pre-rinse

>SMPL sampl i ng oper at i on

aspi r . r at e max mL/ mi n

Aspiration rate for sample transfer

0.001…166.66,

max


= highest possible aspiration rate (depends on type of
dosing unit), select with <CLEAR> key

>SMPL sampl i ng oper at i on

dosi ng r at e max mL/ mi n

Dosing rate for sample transfer

0.001…166.66,

max


= highest possible dosing rate (depends on type of
dosing unit), select with <CLEAR> key

>SMPL sampl i ng oper at i on

sol v. f i l l max mL/ mi n

Filling rate for dilution of sample

0.001…166.66,

max


= highest possible filling rate (depends on type of
dosing unit), select with <CLEAR> key

>SMPL sampl i ng oper at i on

ai r gap l engt h 10 mm

Length of separating bubble in aspiration tub-
ing

1…100


par amet er s

>SMPL meas. cel l r i nsi ng

Settings for rinsing the measuring cell

Navigate with <ENTER>

>SMPL meas. cel l r i nsi ng

ext er nal pump: of f

External pumps for aspirating and rinsing the
measuring cell

on, off

>SMPL meas. cel l r i nsi ng

r i nsi ng r epet i t i ons 3

Number of aspiration and rinsing processes

1…10 recommended: twice

>SMPL meas. cel l r i nsi ng

r i nse t i me 00: 01: 00

Rinsing time with external pump (solvent addi-
tion)

HH:MM:SS

only for 'external pump on'

correction of input with <CLEAR>

>SMPL meas. cel l r i nsi ng

aspi r . t i me 00: 01: 00

Length of aspiration with external pump

HH:MM:SS

only for 'external pump on'

correction of input with <CLEAR>

>SMPL meas. cel l r i nsi ng

aspi r . vol ume 10 mL

Aspiration volume of cell

0.1…500

with 'external pump off'


>SMPL meas. cel l r i nsi ng

aspi r . r at e max mL/ mi n

Aspiration rate for aspirating the cell

0.001…166.66,

max

with 'external pump off'


= highest possible aspiration rate (depends on type of
dosing unit), select with <CLEAR> key

Maximum values for

dosing and filling rates,

etc.

Cylinder vol. Max. rate
2 mL

6.66 mL/min

5 mL

16.66 mL/min

10 mL

33.33 mL/min

20 mL

66.66 mL/min

50 mL 166.66 mL/min
These are pure piston
stroke rates. Filling times
are not taken into account.

Advertising