2 examples of reports - xdos mode – Metrohm 711 Liquino User Manual

Page 87

Advertising
background image

4. Detailed description

711 Liquino, Instructions for use

81

4.4.2

Examples of reports - XDOS mode

Temperature-controlled dosing with volume and time selection

Parameter Report

711 Liquino 0220 711.0011

date 1998-09-30 time 11:51

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: XDOS user method: TempDos1

lot no. 12-2

parameters

>XDOS parameter

XDOS type: volume&time

volume 10.0 mL

dosing time 00:02:00

temp. measuring: on

rec. interval 00:00:10

fill rate max mL/min

auto fill: on

>XDOS dosing units

setup: single

dosing drive: DOS1

dos.unit: D3

port use: custom

dosing port 1

filling port 2

waste port 1

>XDOS monitoring

temperature: on

low limit 20.0 °C

upper limit 30.0 °C

action: wait

alarm pins: none

volume: off

>XDOS preselections

req.ident: id1

>>XDOS prompts

id1 lot no.

id2 id2

id3 id3

start delay 00:00:00

>XDOS reports

temp. auto scale: on

report1: result

report2: curve

report3: comb. list

report4: off

report5: off

report6: off

report7: off

report8: off

------------


← instrument number and program version

← date and time

← 3-line report header

← mode and method name

← ID1, identification with preselected display text

← dosing parameters





← dosing unit settings





← monitoring function settings




← request settings




← report definitions






← identification of reproduced reports

Result Report

711 Liquino 0220 711.0011

date 1998-09-30 time 11:39

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: XDOS user method: TempDos1

lot no. 12-2

total dosing time = 000:02:17

total volume = 10.000 mL

dosing rate = auto

============


← see above




← effective time *

← dosed volume

← dosing rate

← identification of original report

*

total time

including waiting times during infringements of temperature limits or during filling

Advertising