13 check memory card interface, 14 check dosino interface – Metrohm 711 Liquino User Manual

Page 163

Advertising
background image

6. Appendix

711 Liquino, Instructions for use

157

6.5.13

Check memory card interface

This diagnosis step allows the functioning of the internal memory card in-

terface to be checked. The extent of the check depends on the size of
the memory card; this is required for the check to be carried out. With a
128kB memory card the interface will only be checked to the extent of
128kB, even though the interface can address a much larger memory.

Provided that the memory card remains inserted during the whole test proce-
dure and that the mains switch is not operated the test is non-destructive, i.e.
the memory card contents will not be altered by the test.

The memory card is not checked in this test. If it is thought that there is a fault
on the memory card then the test described under section 5.4.12 should be
carried out instead of this test. That test provides information about whether
the memory card control contained in the 711 Liquino and the memory card
itself are functioning properly.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>m. car d i nt er f ace t est

<ENTER>

>m. car d i nt er f ace t est

I NSERT MEMORY CARD

Insert memory card.

Press <ENTER>.

>m. car d i nt er f ace t est

128K ADDRESSABLE

With an SRAM card with 128kB

128K ADDRESSABLE

must appear in the display.

The test is exited by pressing the <STOP> key.

di agnosi s

>dosi no t est


6.5.14

Check Dosino interface

This diagnosis step allows the functioning of the 4 interfaces for controlling

the Dosinos to be checked.

Prepare instrument for diagnosis (see Sect. 6.5.3).

If necessary press key <9> several times until

di agnosi s

>dosi no t est

Advertising