4.9 GLP mode - validation

711 Liquino, Instructions for use

126

Detail Report

711 Liquino 0220 711.0011

date 1998-10-03 time 11:32 - 11:35

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: GLP user method: GLP test

operator D. Miller

drive id Dosino1

dos.unit id Dos20_1

dosing unit size 20 mL

temperature 22.50 °C

liquid water

liquid density 0.9977 g/mL

air density 0.0012 g/mL

dens. std. weights 8.0000 g/mL

calc. factor 1.0034033

nom.v. mass act.vol abs.err rel.err

(mL) (g) (mL) (uL) (%)

----------------------------------------

2.000 1.9900 1.9968 -3.2 -0.161

3.532 3.5100 3.5219 -10.1 -0.285

5.018 4.9900 5.0070 -11.0 -0.220

8.764 8.7300 8.7597 -4.3 -0.049

9.714 9.6700 9.7029 -11.1 -0.114

11.258 11.2200 11.2582 0.2 0.002

13.758 13.7100 13.7567 -1.3 -0.010

16.290 16.2300 16.2852 -4.8 -0.029

17.642 17.6000 17.6599 17.9 0.101

20.000 19.9300 19.9978 -2.2 -0.011

----------------------------------------

>evaluation

intercept (<30) -11.2 uL ok

slope (+/-0.003) 1.00076 ok

error max.vol. (<60) -2 uL ok

-> GLP test PASSED!

============


← instrument number and program version

← date and time

← 3-line report header

← mode and method name

← request texts

← temperature of liquid

← density of liquid at given temperature

← density of air at given temperature *

← density of calibration weight *

← correction factor for taking buoyancy of air into account

when weighing in sample *


← table of values








← results (with limits)


← result



* The correction factor for buoyancy of the air is automatically multiplied with the sample weight.

The factor is calculated as follows:

Correction factor = 1 / ρ

F

( 1 + ρ

L

/ρ

F

ρ

F

/ρ

S

)


where

ρ

F

density of liquid in g/mL

ρ

L

density of air in g/mL

ρ

S

density of balance calibration weight in g/mL


Advertising