Metrohm 711 Liquino User Manual

Page 19

Advertising
background image

2. Installation

711 Liquino, Instructions for use

13


Switch sample changer to next sample

*)

>SEQ sequence st ep 14 ( CONTROL command)

pat t er n 1* * * - - - -

switches sample changer 1 sample further

(start pulse)

>SEQ sequence st ep 15 ( Pause command)

pause 00: 00: 01

wait 1 s

>SEQ sequence st ep 16 ( CONTROL command)

pat t er n 0* * * - - - -

resets remote interface

(end pulse)

>SEQ sequence st ep 17 ( SCAN command)

pat t er n * * * 1* - - -

wait for ready message from sample

changer


711 Liquino – 730 Sample Changer – Titrino

730 Sample Changer with Titrino and 711 Liquino for automated pipetting of
different samples on the sample changer into a special beaker on the sam-
ple rack.

711

B

A

Titrino

C

D

730




The 711 Liquino can be used in any automated system for pipetting tasks
(PIP mode). In this case the 730 Sample Changer controls the working pro-
cess. The 711 Liquino is used in the PIP mode. The determination of the
samples is carried out by the Titrino which, like the Liquino, is started from
the sample changer.
This arrangement of instruments requires a special remote cable
(6.9921.165). Other special cables for similar instrument combinations are
available on request.

*)

For this function the sample changer must be appropriately programmed with Scan and Control com-
mands!

cable 6.9921.165

Command to
730 Sample Changer

SCN:Rm: ready1 ←






CTL:Rm: START device1→

Advertising