4.4 XDOS mode - controlled dosing

711 Liquino, Instructions for use

82

Curve Report

711 Liquino 0220 711.0011

date 1998-09-30 time 11:39

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: XDOS user method: TempDos1

lot no. 12-2

============









← volume/temperature curve







Combined List

711 Liquino 0220 711.0011

date 1998-09-30 time 11:39

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: XDOS user method: TempDos1

lot no. 12-2

time vol.(mL) temp(°C) event

----------------------------------------

000:00:00 0.000 *** START ***

000:00:00 0.000 27.5

000:00:10 0.828 28.5

000:00:20 1.656 29.5

000:00:25 2.075 T-limit,wait

000:00:30 2.075 30.0 upper T-limit

000:00:40 2.075 30.0 upper T-limit

000:00:42 2.075 continue

000:00:50 2.751 29.5

000:01:00 3.584 29.5

000:01:10 4.416 29.5

000:01:20 5.249 29.0

000:01:30 6.081 29.5

000:01:40 6.914 29.5

000:01:50 7.746 29.0

000:02:00 8.579 29.0

000:02:10 9.411 28.5

000:02:17 10.000 *** END ***

000:02:20 28.5

000:02:30 28.5

============









← tabular list of volumes, temperatures and
events

← start of dosing

← temperature limit reached, dosing interrupted

← temperature back in limit, dosing continued







← end of dosing

← procedure ended after dosing unit has been filled.


Temperature curve

Volume curve

Temperature limit

Temperature
limit reached

Dosing interrupted