Defining dosing units


In order for the complete tubing system of a Dosino to be fully filled with the
reagent solution without any air bubbles the 711 Liquino must know the di-
mensions of the connected tubing. For this reason a dosing unit can be de-
fined in the configuration menu and then selected for the method.

*)


Except for the "Sampling dosing unit" in the SMPL mode the standard port
occupancy for all modes is as follows:

Port 1:

dosing unit output

Port 2 (bottle):

refilling opening


CONFIG

A

Press <CONFIG> and then 3x <>.

Di spl ay:

conf i gur at i on

>dosi ng uni t s

ENTER

YES

Access the submenu ‘dosing units’ with <ENTER>.

Di spl ay:

> dosi ng uni t s

dos. uni t : D1

DOS NR

SELECT

P

One of the ten configurable dosing units can be selected with the

<SELECT> key. Press <ENTER> to access the settings for the corre-
sponding dosing unit. The index for the dosing unit is located in the top
right-hand corner of the display.

Di spl ay:

>dosi ng uni t s 1

I D D1

Enter an identification text for the dosing unit or leave the preset entry

as it is. Text input is explained on page 63.

Accept the entry with <ENTER>.

Di spl ay:

>dosi ng uni t s 1

l engt h t ube1 400 mm

Now enter the length of tubing at Dosino port 1 (for tandem dosing

including the tubing length up to the T-piece) and accept the entry with
<ENTER>.

Di spl ay:

>dosi ng uni t s 1

di am. t ube1 2. 0 mm

The inner diameter of the tubing at Dosino port 1 can be entered here

and confirmed with <ENTER>.

Di spl ay:

>dosi ng uni t s 1

l engt h t ube2 250 mm

Now enter the length of the tubing at Dosino port 2 (bottle) and accept

the entry with <ENTER>.

Di spl ay:

>dosi ng uni t s 1

*) In a single method ‘standard’ can also be selected for the dosing unit used. For the modes XDOS, CONT
and GLP ‘standard’ means:

Port1 (dosing port):

L = 40cm, D = 2mm

Port2 (bottle):

L = 25cm, D = 2mm.

For PIP ‘standard’ means:

Port1 (pipetting tubing):

L = 150cm, D = 2mm

Port2 (bottle):

L = 25cm, D = 2mm.

For SMPL ‘standard’ means:

Port1 (sample tubing):

L = 40cm, D = 2mm

Port2 (measuring cell):

L = 16cm, D = 2mm

Port3 (rinsing agent):

L = 50cm, D = 2mm.