3 Introduction

711 Liquino, Instructions for use

59






<ENTER>



<ENTER>






<SELECT>











<ENTER>



2x <QUIT>

Di spl ay:

>GLP opt i ons

bal ance- I d

Enter an unambiguous identification for the balance used (e.g. serial

number). The input of text is explained on 63.

Di spl ay:

par amet er s

>GLP r epor t s

As the ‘GLP options’ submenu contains no further menu items, an

automatic return is made to the submenu selection. Press <ENTER>
to open the submenu ‘GLP reports’. The reports which are to be print-
ed out automatically at the end of the GLP test are defined here.

Di spl ay:

>GLP r epor t s no. 1

r epor t : of f

The <SELECT> key can be used to select from the following reports:

‘off’: no report

‘result’: summary of the results

‘parameter’: parameter settings

‘curve’: graph of the measurement

‘detail’: results in detail with list of measuring points

‘ff’: page feed (may be useful between 2 reports)

Select the reports to be printed out and press the <ENTER> key after
each selection. Up to 5 entries can be made.

Press <QUIT> 2x to exit the parameter menu.

All the necessary parameters have now been entered!



Various GLP methods can be edited and stored in the method memory.
Each of these memories can then be reloaded by pressing a few keys.
The method memory is described on page 128.