Metrohm 711 Liquino User Manual

Page 98

Advertising
background image

4.5 SEQ mode - complex processes

711 Liquino, Instructions for use

92

>SEQ sequence st ep x

pat t er n * * * * * - - -

Bit pattern of input lines of the remote
socket

00000---…11111---

only for 'interface: Remote'


<CLEAR> can be used to delete the entry.

>SEQ sequence st ep x

st r i ng

Character string in input buffer of
serial RS232 interface

15 ASCII characters

only for 'interface: RS'


Input in text input mode. Press
<ALPHA> key.


<CLEAR> can be used to delete the entry.


CONTROL


>SEQ sequence st ep x

command: CONTROL

Operate interface

Edit parameters with <ENTER>


>SEQ sequence st ep x

i nt er f ace: Remot e

Select interface with

<SELECT>

Remote

RS

= output lines of remote socket
= send data via RS232 interface

>SEQ sequence st ep x

pat t er n * * * * - - - -

Bit pattern for setting output lines of
remote socket

0000----…1111----

only for 'interface: Remote'


<CLEAR> can be used to delete the entry.

>SEQ sequence st ep x

st r i ng

Character string which is to be sent
via the RS232 interface

15 ASCII characters

only for 'interface: RS'


Input in text input mode. Press
<ALPHA> key.


<CLEAR> can be used to delete the entry.


ENDSEQ


>SEQ sequence st ep x

command: ENDSEQ

End mark of the sequence

This command can be used for marking the end
within a sequence. It is therefore very useful for
testing long sequences.
It is not necessary to set an ENDSEQ mark at the
end of a sequence.

1 = line active (low level)
0 = line inactive (high level)
* = condition will not be changed

1 = line active (low level)
0 = line inactive (high level)
* = condition will not be tested