4.5 SEQ mode - complex processes
711 Liquino, Instructions for use
92
>SEQ sequence st ep x
pat t er n * * * * * - - -
Bit pattern of input lines of the remote
socket
00000---…11111---
only for 'interface: Remote'
<CLEAR> can be used to delete the entry.
>SEQ sequence st ep x
st r i ng
Character string in input buffer of
serial RS232 interface
15 ASCII characters
only for 'interface: RS'
Input in text input mode. Press
<ALPHA> key.
<CLEAR> can be used to delete the entry.
CONTROL
>SEQ sequence st ep x
command: CONTROL
Operate interface
Edit parameters with <ENTER>
>SEQ sequence st ep x
i nt er f ace: Remot e
Select interface with
<SELECT>
Remote
RS
= output lines of remote socket
= send data via RS232 interface
>SEQ sequence st ep x
pat t er n * * * * - - - -
Bit pattern for setting output lines of
remote socket
0000----…1111----
only for 'interface: Remote'
<CLEAR> can be used to delete the entry.
>SEQ sequence st ep x
st r i ng
Character string which is to be sent
via the RS232 interface
15 ASCII characters
only for 'interface: RS'
Input in text input mode. Press
<ALPHA> key.
<CLEAR> can be used to delete the entry.
ENDSEQ
>SEQ sequence st ep x
command: ENDSEQ
End mark of the sequence
This command can be used for marking the end
within a sequence. It is therefore very useful for
testing long sequences.
It is not necessary to set an ENDSEQ mark at the
end of a sequence.
1 = line active (low level)
0 = line inactive (high level)
* = condition will not be changed
1 = line active (low level)
0 = line inactive (high level)
* = condition will not be tested