3 Introduction

711 Liquino, Instructions for use

49

<SELECT>

<ENTER>


















<ENTER>

2x <QUIT>

In this menu item <SELECT> can be used to choose whether a tem-
perature curve should be scaled automatically or manually. Accept the
default value ‘on’ with <ENTER>.

Di spl ay:

>XDOS r epor t s no. 1

r epor t : of f

The <SELECT> key can be used to select from the following reports:

‘off’: no report

‘results’: summary of the results

‘parameters’: parameter settings

‘curve’: graph of the measurements and additions

‘V-list’: list of volume measuring points

‘T-list’: list of temperature measuring points

‘comb. list': combined volume/temperature measuring point list

‘ff’: page feed (may be useful between 2 reports)

Select the reports to be printed out and press the <ENTER> key after
each selection. Up to 8 entries can be made.
Press <QUIT> 2x to exit the parameter menu.
All the necessary parameters have now been entered!



Various XDOS methods can be edited and stored in the method memory.
Each of these methods can then be reloaded by pressing a few keys. The
method memory is described on page 128.